This is explanation on Error 6 from Service Manual:
EEROM (A1U5) meter configuration corrupted or EEROM not initialized. At power up, A1U6 reads CRC from EEROM (A1U5) location, then reads configuration data from another location in EEROM, calculates an algorithm to compare to the CRC. If an error is determined, the meter configuration is set to factory default and Error 6 is displayed.
You can find service manual with troubleshooting procedure here: http://www.fluke.com/Fluke/usen/Support/Manuals/default.htm?ProductId=56082
615 views
Usually answered in minutes!
the machine needs to be calibrated
I thinks it's faulty - not sure that calibration will fix it, it needs repair first.
×