Invacare Supply Group Health & Beauty - Popular Questions, Answers, Tips & Manuals


You should call the manufacturer for this type answer you need.

Invacare Supply... | 30 views | 0 helpful votes


Hi flashing lights 8 times (Controller fault.). if the batteries haven't been used for several years U might have to replace them .U can buy a BATTERY LOAD TESTER OF E/BAY cost approx,$36..this will give U an accurate test on the condition of the batteries ,good luck
26346437-e5jkhq4oawb2rbtt3t2s2xlr-3-0.jpg

Invacare Supply... | 151 views | 1 helpful votes


Make sure that you are charging the chair 12-14 hrs before driving. charge with the joystick on, make sure both freewheel levers are in locked position.

I’m happy to assist further over the phone at https://www.6ya.com/expert/bill_bcbb89783b08c489

Invacare Supply... | 131 views | 0 helpful votes


Test the battery voltage at the joystick where a programmer or charger would plug in. There are 3 holes, the top or number one hole is the positive hole for a voltage meter to plug into and the bottom or number two hole is the negative hole for the volt meter. Plug in the meter and see what the voltage is when sitting still and keep the meter in and see what the voltage does while you are driving the chair. The voltage should not drop more than 1 or 2 volts from that which you had when the chair was not moving. If the voltage drops more than that then there is a problem with charging or the new batteries. If the voltage remains good then the joystick or speed knob control could be the problem.

Invacare Supply... | 292 views | 1 helpful votes


hi, yes i can, i repaired one last week,

please contact me my details are on the website

Regards
gewizrepairs.co.uk

Invacare Supply... | 330 views | 0 helpful votes


The red light is the power light and if the red light is blinking the charger is bad or the wiring is bad on the chair or the batteries are bad. Usually it's the charger.

Invacare Supply... | 758 views | 0 helpful votes


If you can ,first check for fluid loss, if this doesn't solve the problem you my have a bad drive bearing. GOOD LUCK FRIEND.

Invacare Supply... | 147 views | 0 helpful votes


It is usually on the serial number sticker, which can be hard to find or have been removed/fallen off. If you can find the serial number, your vendor or the Invacare Co. can tell you the model, etc.

Invacare Supply... | 136 views | 0 helpful votes


You are getting a motor error code. Check motor wiring and connection to the main control box. Motor may have gotten unplugged when changing the batteries.

Invacare Supply... | 296 views | 0 helpful votes


The red light on the transformer/Charger means the charger is bad. Sometimes it can also be battery harness wiring. Most often it is the charger that needs replaced.

Invacare Supply... | 307 views | 0 helpful votes


The M91 has an onboard charger. When unplugging the charger power cord from the wall outlet does not stop the chair from being in the charge mode, The charger has to be unplugged from the wheelchair wiring harness or main control module under the plastic body parts. The on-board chargers fail like this sometimes.

Invacare Supply... | 287 views | 0 helpful votes


The chair may be in what they call SLOW MODE or the batteries are weak. If you have powered seating on the chair the seat has sensors that make the chair drive slow or not drive at all when the seat is tilted or reclined or raised. A sensor can go bad or get out of position. The joystick could have different programming for different drive modes. Try changing drive modes.

Invacare Supply... | 205 views | 0 helpful votes


They are diagnostic codes to help troubleshoot a possible problem with your wheelchair. The lights blink in patterns. You have to count the number of blinks and what color they are.

Invacare Supply... | 192 views | 0 helpful votes


The ON button is at the top front of the joystick. If this button does not turn on the red, yellow and green lights there are many possibilities as to why it is not turning on. Batteries have to be checked first for proper voltage. Then wiring needs to be checked for proper connections and fuses checked.

Invacare Supply... | 86 views | 0 helpful votes


1 red + 5 green is left parking brake fault. This is a pretty generic code and can be something as simple as the brake/clutch lever is in free-wheel, or it can mean something more significantly wrong, depending on how it is wired in this model. (We don't work with Prontos a lot here.)

Invacare Supply... | 179 views | 0 helpful votes


http://www.invacare.com/doc_files/1143190.pdf
Owners manual for TDX SP link above.

First check if you have power present at the charging port. If you have power (24 volts) then the batteries and wiring are not the issue. If you don't then check the battery post fuses. These are "L" shaped brackets with a white square fuse held on the positive post with a through bolt and nut. On older chairs these are known to become loose and or blow just when you need power or stop suddenly.
Also check the big Anderson connectors at the power point at the end of the battery box. Check for solid connections. If there is 24 volts here to the control box, next check the controller connector.
If there is +24 at the controller connector but 0 at the joystick charger port you have big issues. Either the traces connectors in the controller are fried or your joystick has major issues.

Invacare Supply... | 439 views | 0 helpful votes

Not finding what you are looking for?
Invacare Supply Group Logo

218 questions posted

Ask a Question

Usually answered in minutes!

Top Invacare Supply Group Health & Beauty Experts

Rob Hill
Rob Hill

Level 3 Expert

1414 Answers

Marvin
Marvin

Level 3 Expert

85239 Answers

Lawrence Oravetz

Level 3 Expert

10548 Answers

Are you an Invacare Supply Group Health and Beauty Expert? Answer questions, earn points and help others

Answer questions

Manuals & User Guides

Loading...