Hi library336
The
Omron
HEM-780 generally has 5 Error Messages:
1]ERROR MESSAGE: A single (E) Error message indicates movement during the measurement process, resulting in an inconclusive reading or the absence of a Blood Pressure or Pulse reading caused by a weak Arterial Pulse.
SOLUTION: Remove the cuff. Wait 10-15 minutes. Repeat measurement using approximately 40mmHg more pressure than in the previous measurement.
2]ERROR MESSAGE: A Double (EE) Error message indicates an under-inflated cuff. Remove the cuff. Wait 2-3 minutes.
SOLUTION: Repeat measurement using approximately 40mmHg more pressure than in the previous, under-inflated measurement.
3]ERROR MESSAGE: An E directly below another E Error message indicates an over-inflated cuff (in excess of 300 mmHg). Remove the cuff.
SOLUTION: Wait 2-3 minutes. Repeat measurement using approximately less pressure than in the previous, over-inflated measurement.
4]ERROR: A Blank Screen indicates low or depleted batteries.
SOLUTION: Remove and replace all 4 AA Batteries. Do not mix used and new batteries.
5]ERROR MESSAGE: An (Er) Error Message indicates an abnormal memory function caused by a faulty internal component.
SOLUTION:
Call Omron Customer
Service, Toll Free, on (in the USA) 1-800-634-4850 (or on the equivalent Toll Free or Freefone Number in your country).
I hope this helps.
If this Answer goes some way towards helping to resolve
your Query, Problem or Fault, please take the time to Rate it.
Doing so helps Solution
Providers, like me, better understand the needs and the point of view of
Askers, like you, and guides us in exploring the best ways to provide the most
comprehensive, useful and high quality Responses and Solutions we can.
If you need further
help, please do not hesitate to post another Question or add a Comment to this
Question.
Thanks for using FixYa.
BATTERIES HAVE BEEN REPLACED NEW ONE,STILL THE SAME PROBLEM OCCURS.CAN U GIVE ANTHER SOLUTION PLEASE?
Blood pressure pumps up to 40 back to 30 then shows errir
×