Mitutoyo error code E940?

Mitutoyo cmm error e940

Posted by Anonymous on

Ad

1 Answer

Angel Garcia

  • Level 1:

    An expert who has achieved level 1.

    MVP:

    An expert that got 5 achievements.

    Habit-Forming:

    Visited the website for 3 consecutive days.

    Scholar:

    An expert who has written 20 answers of more than 400 characters.

  • Contributor
  • 23 Answers

Mitutoyo is a brand that produces precision measuring instruments such as micrometers, calipers, and height gauges. An error code E940 on a Mitutoyo instrument typically indicates a problem with the instrument's communication interface. This can be caused by a variety of issues, such as a problem with the instrument's cable or software, or a problem with the computer or device to which the instrument is connected.
To troubleshoot the issue, you should first check the connections between the instrument and the computer or device to ensure that they are properly connected and secure. If the connections appear to be fine, you may need to update the software on the instrument and computer or device to resolve the issue.

Posted on Jan 12, 2023

Ad

Add Your Answer

×

Uploading: 0%

my-video-file.mp4

Complete. Click "Add" to insert your video. Add

×

Loading...
Loading...

Related Questions:

0helpful
1answer
1helpful
1answer
0helpful
1answer

What mean error E812 on Mitutoyo BHN 305

I think you may have to contact their technical support for this answer. Such a link is on the following page: https://www.mitutoyo.com/products-and-solutions/software/mcosmos/
0helpful
1answer
0helpful
1answer

What does Mitutoyo error code E380 mean?

These 2 signals are drive prohibition (overtravel) inputs in forward and reverse directions. If Drive Prohibition Input Setting (Pn504) is 1, you can use the setting of Stop Selection (Pn505) to select the operation to be taken upon input of each prohibit signal. If Drive Prohibition Input Setting (Pn504) is 2, drive prohibition input protection (E380) actuates upon input of a drive prohibition.
1helpful
2answers

How to correct an e671 error message

I realise this question is old, but you never know, might help someone in the future. This particular error is that the probe is not correctly seated, usually the magnets have become slightly loose. The solution is to take the probe head off the cmm, and push the magnets against your surface table or suitable flat surface, this should push the magnets into their correct position. If this doesn't help, it may be that you've managed to break that particular part, if you have a probe rack, try a different probe. Lastly, if you crashed the CMM by hitting something, ensure you didn't bend your probe in the process (and the ruby is still there!), this doesn't help the error code fix, but worth checking as the cmm crashing into something at speed is the most likely cause of this E671 error.
0helpful
1answer

Philips media converter wont work it says my device is not connected and conversion wont start at all

For some reason, if you have Adobe's CMM installed on your computer, the Media Converter for Philips is unable to start up and will error out every time. You have to uninstall Adobe's CMM (any version), once you do that, the Media Converter for Philips by ArcSoft will work as it should. I found this conflict when I viewed the error in Windows Event Viewer where it listed that the conflict was being caused by the Adobe CMM.dll, I uninstalled Adobe's CMM from my system and the media converter works fine now.
3helpful
2answers

I have an error message: E-940. Can anyone help???

E940 - Main Switch Error

The detection mechanism is as follows:
· The machine reaches the auto-off time.
· The solenoid mounted inside the main switch turns on for 3 seconds.
· If the main switch does not turn off, the solenoid is turned off for 1
seconds.
· The solenoid is once again turned on for 3 seconds.
· If the main switch does not turn off at this point, the solenoid turns off
and E940 lights.
Possible Causes -
The main switch is defective.
The main control board is defective.
1. Turn on the main switch.
2. Enter SP5-810 (SC code reset) and press the Enter key.
3. Exit SP mode and turn the main switch off and on.
Not finding what you are looking for?

29 views

Ask a Question

Usually answered in minutes!

Top Mitutoyo Measuring Tools & Sensors Experts

Brad Brown

Level 3 Expert

19166 Answers

vince

Level 3 Expert

1543 Answers

ADMIN Andrew
ADMIN Andrew

Level 3 Expert

55586 Answers

Are you a Mitutoyo Measuring Tool and Sensor Expert? Answer questions, earn points and help others

Answer questions

Manuals & User Guides

Loading...