Question about By HoMedics Automatic Wrist Blood Pressure Monitor 1ea
...blood pressure value is displayed when an excessively low (or high) reading is detected OR as a result of one of the following: 1. Cuff has been placed incorrectly. Solution: Wrap cuff properly so ...
Question about By Homedics Automatic Blood Pressure Monitor 1ea
...blood pressure she gets error 3 every time. When I do it, it works. What could we be doing wrong. 2 things could be causing the problem, when your wife is doing it. 1. Improper placement of the cuff. ...
Question about By HoMedics Automatic Wrist Blood Pressure Monitor 1ea
blood pressure monitor reads E- error. What is the reason? Not keeping the wrist still? It depends on which E-error you're getting. EE - indicates several things: Cuff has been placed incorrectly.
...error E1 indicates an air circuit abnormality. Wait one minute and try inflating the cuff again. If the problem continues, contact HoMedics' Customer Support 1-800-466-3342. Hope this helped and ...
monitor the E1 error code indicates an "Air circuit abnormality". Solution: Measure again. If this continues to occur, contact HoMedics Customer Support 1-800-466-3342. Hope this helped you and thanks
Question about By HoMedics Automatic Wrist Blood Pressure Monitor 1ea
...error code. 1. Cuff was placed incorrectly. Wrap cuff snugly around your wrist 1" above the heel of your left hand. 2. Moving or talking during measurement. Measure again keeping your wrist steady ...