Hi,
One of the cause for this problem is the faulty backup battery. normally ecg machines runs on battery even if the power is connected. The power supply of ecg machines is meant for charging purpose only and doesn't have that much power to run the machine. also the companies will not recommend to take ecg at mines on condition.
First of all check the charging voltage is there or not. The battery charging voltage is 13.8v . check is the voltage is there at the pins of battery ,in mains on condition. If its ok the the battery have to be replaced. Normally the battery will be faulty.
ok
778 views