Question about Schiller Cardiovit At 1 Ecg Machine

1 Answer

The schiller at-1 ran out of paper, I put some more in. Now it will not print any suggetions?

Posted by on

  • Electro Med Services
    Electro Med Services Mar 18, 2013

    HI, check this link for know more about schiller AT 1 Problems and solutions and setting of A1 http://electro-medical.blogspot.in/2011/...

×

Ad

1 Answer

  • Level 3:

    An expert who has achieved level 3 by getting 1000 points

    Superstar:

    An expert that got 20 achievements.

    All-Star:

    An expert that got 10 achievements.

    MVP:

    An expert that got 5 achievements.

  • Schiller Master
  • 6,694 Answers

HI,
check the paper u have inserted. it should be Schiller original paper with page marking. also clean the paper sensor. u can see the paper sensor near the side of printing head. a small square portion. clean it with an ear bud, don't use any solution for cleaning only ear bud.
ok

Posted on Jan 04, 2011

Ad

1 Suggested Answer

6ya6ya
  • 2 Answers

SOURCE: I have freestanding Series 8 dishwasher. Lately during the filling cycle water hammer is occurring. How can this be resolved

Hi,
a 6ya expert can help you resolve that issue over the phone in a minute or two.
best thing about this new service is that you are never placed on hold and get to talk to real repairmen in the US.
the service is completely free and covers almost anything you can think of (from cars to computers, handyman, and even drones).
click here to download the app (for users in the US for now) and get all the help you need.
goodluck!

Posted on Jan 02, 2017

Ad

Add Your Answer

Uploading: 0%

my-video-file.mp4

Complete. Click "Add" to insert your video. Add

×

Loading...
Loading...

Related Questions:

2 Answers

When auto print is pressed the machine blanks out and nothing prints


Try this link to slide share for a colour rendition of the user manual
http://www.slideshare.net/DiscountCardiology/schiller-at-2-ekg-manual

Aug 30, 2012 | Schiller Cardiovit At 2 Light Ecg Machine

2 Answers

Just ordered a schiller at1 ekg machine, just realized it is 3 channel, does it have a 1 channel mode where just one tracing will print?


hi,
u can make it single channel one by changing its setup. it will print single trace in the same paper. when u select this option it use more paper than the three trace.
ok

Dec 06, 2010 | Schiller Cardiovit At 1 Ecg Machine

2 Answers

The schiller At-2 we have in the office is acting up. Model number 021-02609. Office number is 480-889-0508. The paper feeds and doesn't stop were needed and the ECG being preformed on a patient are not...


HI,
check the paper u have inserted. it should be Schiller original paper with page marking. also clean the paper sensor. u can see the paper sensor near the side of printing head. a small square portion. clean it with an ear bud, don't use any solution for cleaning only ear bud.
ok

Oct 13, 2010 | Schiller Cardiovit At 2 Light Ecg Machine

3 Answers

Schiller AT-2 Plus, placed new paper in machine and machine keeps stating "out of paper" and will not print unless I hold down the manual print.


Your machine has the wrong paper in it or the paper sensor is not working properly.  There should be a black mark on the top left hand corner of each page.  The correct ekg paper and ekg repair is available from Medical Device Depot.

Feb 15, 2010 | Schiller Welch Allyn Cardiovit At 2...

2 Answers

Schiller AT-1


check paper alignment
maybe papermark sensor problem also

Oct 16, 2009 | Schiller Cardiovit At 1 Ecg Machine

2 Answers

The schiller at 2 plus ecg machine not printing ecg


Does it work in auto mode ? are you putting paper correctly means , there is a black mark on paper which is sensed by paper sensor .If you put paper up side down then paper sensor won't be able to sense any paper & gives you error for rifilling paper .If this doesn't help you then You need to replace battrey . Thanks

Jul 15, 2009 | Schiller Welch Allyn Cardiovit At 2...

2 Answers

Schiller machine is not printing


If the unit print lightly or not printing anything at all?
The most common causes for not printing ECGs is one of those:
1- Too little heat on the stylus
2- Insufficient stylus pressure on the paper
Solutions:
1- Using an insulated probe, such as a screwdriver, gently press the stylus onto the paper.
2- If a dark line appears on the paper, the problem is too little pressure, but if no dark line appears, the problem is not enough heat.

Jun 19, 2009 | RTO Schiller CARDIOVIT AT-1...

1 Answer

Schiller cardiovit at-1 not printing clearly


hi clean the thermal head of the printer with the tissue paper
Zubair
0091-9849156725

Jun 19, 2009 | RTO Schiller CARDIOVIT AT-1...

Not finding what you are looking for?
Schiller Cardiovit At 1 Ecg Machine Logo

Related Topics:

248 people viewed this question

Ask a Question

Usually answered in minutes!

Top Schiller Medical Experts

Electro Med Services...
Electro Med Services...

Level 3 Expert

6694 Answers

The Knight
The Knight

Level 3 Expert

74243 Answers

Are you a Schiller Medical Expert? Answer questions, earn points and help others

Answer questions

Manuals & User Guides

Loading...