Question about Respironics Inspiration Elite Compressor Nebulizer with SideStream Reusable

1 Answer

The nebulizer is still working and creating mist but the mist isn't coming out very well and it takes forever to get through the medicine.

Posted by on

Ad

1 Answer

  • Level 1:

    An expert who has achieved level 1.

  • Contributor
  • 4 Answers

You need a new mouthpiece/mask. For now, trying boiling the one you have for two minutes.

Posted on Dec 06, 2010

Ad

1 Suggested Answer

6ya6ya
  • 2 Answers

SOURCE: I have freestanding Series 8 dishwasher. Lately during the filling cycle water hammer is occurring. How can this be resolved

Hi,
a 6ya expert can help you resolve that issue over the phone in a minute or two.
best thing about this new service is that you are never placed on hold and get to talk to real repairmen in the US.
the service is completely free and covers almost anything you can think of (from cars to computers, handyman, and even drones).
click here to download the app (for users in the US for now) and get all the help you need.
goodluck!

Posted on Jan 02, 2017

Ad

Add Your Answer

Uploading: 0%

my-video-file.mp4

Complete. Click "Add" to insert your video. Add

×

Loading...
Loading...

Related Questions:

1 Answer

Airial nebulizer does not mist


I am very happy to read your articles it's very useful for me, and I am completely satisfied with your website. All comments and articles are very useful and very good. Your blog is very attention-grabbing. I am loving all of the in. turn you are sharing with each one!.... <a href="http://www.easyfit.in" title="Nebulizer" target="_blank">Nebulizer </a>

65fe434f-3e14-402b-a673-8239a923a51f.jpg


a134c137-3621-45bb-9005-5afd2c88e1eb.jpg

Apr 02, 2013 | Medline Nebulizer Portable Compressor Sys...

1 Answer

No medicine coming out of nebulizer


Replacement of neb kit the most common cause. Try this first. Then get back with me.

May 31, 2010 | Respironics HS800 Disposable Nebulizer Kit...

1 Answer

My nebulizer has air coming out but no mist


The tube from the medicine is probably blocked. Generally needs factory service.

Apr 16, 2010 | Inspiration 626 Compressor Nebulizer...

3 Answers

Omron nebulizer will not mist. All tubing is intact. What to do?


You'll read a lot about making sure that the filter is clean and that the medicine cup is cleaned and that hard water can effect the operation of the nebulizer. However, the problem is most likely in the pump. Even though the pump is running, it is not pushing enough air into the medicine cup. You may even feel air coming out of the tube. When unrestricted, the air will come out, but, once restricted by the medicine cup, the air may not be fast enough to nebulize the liquid. If you want to confirm this, take the tube and hold it to the end of a bike pump, if, while pumpung, you successfully see mist, then you know for sure.

Fixing the pump is not too difficult. Over time, the chamber where the air is compressed becomes dry and, just like a dry bike pump, the air escapes from the chamber due to poor seals. The pump is pretty simple. Make sure it is unplugged from power and remove the screws holding the housing together. Inside you will see the motor, a fan, a power switch and the intake and outtake tubes. Follow the outtake tube to the pump. The junction where the tubes meet the pump is the location of the air chamber. You may need to remove the screws that secure the tubes to the chamber. Once apart, take some grease and lightly coat all of the interior surfaces of the chamber. When finished, reassemble and you're good to go!

Mar 23, 2009 | Omron Healthcare Omron NE-C25 COMP-Air...

2 Answers

Very little mist coming out


You'll read a lot about making sure that the filter is clean and that the medicine cup is cleaned and that hard water can effect the operation of the nebulizer. However, the problem is most likely in the pump. Even though the pump is running, it is not pushing enough air into the medicine cup. You may even feel air coming out of the tube. When unrestricted, the air will come out, but, once restricted by the medicine cup, the air may not be fast enough to nebulize the liquid. If you want to confirm this, take the tube and hold it to the end of a bike pump, if, while pumpung, you successfully see mist, then you know for sure.

Fixing the pump is not too difficult. Over time, the chamber where the air is compressed becomes dry and, just like a dry bike pump, the air escapes from the chamber due to poor seals. The pump is pretty simple. Make sure it is unplugged from power and remove the screws holding the housing together. Inside you will see the motor, a fan, a power switch and the intake and outtake tubes. Follow the outtake tube to the pump. The junction where the tubes meet the pump is the location of the air chamber. You may need to remove the screws that secure the tubes to the chamber. Once apart, take some grease and lightly coat all of the interior surfaces of the chamber. When finished, reassemble and you're good to go!

Jan 25, 2009 | Mabis CompMist Compressor Nebulizer

Not finding what you are looking for?
Respironics Inspiration Elite Compressor Nebulizer with SideStream Reusable Logo

Related Topics:

665 people viewed this question

Ask a Question

Usually answered in minutes!

Top Respironics Health & Beauty Experts

Cindy Wells

Level 3 Expert

5129 Answers

David Payne
David Payne

Level 3 Expert

14161 Answers

 Bob S
Bob S

Level 1 Expert

5 Answers

Are you a Respironics Health and Beauty Expert? Answer questions, earn points and help others

Answer questions

Manuals & User Guides

Loading...