Question about Omron Automatic Blood Pressure Monitor Options Large Adult Cuff Fits Arms 13 To 17

1 Answer

Omron 780 My machine registers readings that are all over the place and incorrect. ie. the reading will be 150/80 and two minutes later might be 90/84 and then 65/100 My blood pressure in the doctor's office is usually about 130/80 so my machine is obviously not working. Is this the callibration? How do I fix it? I haven't tried anything as I don't know what to do. thank you MS

Posted by on

1 Answer

  • Level 3:

    An expert who has achieved level 3 by getting 1000 points

    Top Expert:

    An expert who has finished #1 on the weekly Top 10 Fixya Experts Leaderboard.

    Superstar:

    An expert that got 20 achievements.

    All-Star:

    An expert that got 10 achievements.

  • Omron Master
  • 21,384 Answers

Check the batteries.

Posted on Aug 04, 2017

5 Suggested Answers

6ya6ya
  • 2 Answers

SOURCE: I have freestanding Series 8 dishwasher. Lately during the filling cycle water hammer is occurring. How can this be resolved

Hi,
a 6ya expert can help you resolve that issue over the phone in a minute or two.
best thing about this new service is that you are never placed on hold and get to talk to real repairmen in the US.
the service is completely free and covers almost anything you can think of (from cars to computers, handyman, and even drones).
click here to download the app (for users in the US for now) and get all the help you need.
goodluck!

Posted on Jan 02, 2017

r321567499
  • 74 Answers

SOURCE: inaccurate reading

I'm not sure why you need it calibrated if your Docotor says it's accurate but here is a web site link....

I hope this helps.
http://www.omronhealthcare.com/product/s1-1126-186-manual-monitor

Posted on Dec 25, 2008

Max J
  • 22 Answers

SOURCE: Manual for Omron HEM-780 Automatic Blood Pressure Monitor

Get a manual at :
http://www.omronhealthcare.com/product/manuals.asp

Posted on Sep 07, 2009

  • 6 Answers

SOURCE: Have Omron 780 Blood pressure monitor. 9 times out

It appears you may have a leak in the cuff. This will need to be verified by an authorised service facility. Contact OMRON they should be able to assist you.
CONSUMER SUPPORT TOLL FREE NUMBER: 1-877-216-1333

Customer Service Representatives are available
Monday through Friday - 8:30am to 4:30pm CT

Posted on May 07, 2010

  • 4736 Answers

SOURCE: My Omron HEM-780 needs to

Recalibration normally has to be done by the manufacturer. As they have to certify the calibration. Keep in mind, that it is not unusual for your BP to read a bit higher at the doctor's office, than it does at home. It's sarcastically called "White coat hypertension." Which translated means, your blood pressure rises a bit at the doctor's office due to anxiety. Or it's being taken at a different time of day, than you take it at home, in a relaxed atmosphere.

Hope this helped you.

Posted on May 16, 2011

Add Your Answer

Uploading: 0%

my-video-file.mp4

Complete. Click "Add" to insert your video. Add

×

Loading...
Loading...

Related Questions:

1 Answer

Blood pressure monitor readings too low, inaccurate


I have used Omron monitors for around 2 years now and I have not found any problem with it. The error may be because you went to the doctor, your blood pressure must have changed since there was a time gap. There is no need to get worried.

Sep 05, 2012 | Omron Latest Model HEM-780 Automatic Blood...

1 Answer

I just put new batteries in my Omron HEM-780. Doesn't work. When it finishes deflating it always says E7. What's going on?


Hi library336

The Omron HEM-780 generally has 5 Error Messages:


1]
ERROR MESSAGE: A single (E) Error message indicates movement during the measurement process, resulting in an inconclusive reading or the absence of a Blood Pressure or Pulse reading caused by a weak Arterial Pulse.

SOLUTION: Remove the cuff. Wait 10-15 minutes. Repeat measurement using approximately 40mmHg more pressure than in the previous measurement.

2]
ERROR MESSAGE: A Double (EE) Error message indicates an under-inflated cuff. Remove the cuff. Wait 2-3 minutes.

SOLUTION: Repeat measurement using approximately 40mmHg more pressure than in the previous, under-inflated measurement.

3]
ERROR MESSAGE: An E directly below another E Error message indicates an over-inflated cuff (in excess of 300 mmHg). Remove the cuff.

SOLUTION: Wait 2-3 minutes. Repeat measurement using approximately less pressure than in the previous, over-inflated measurement.


4]
ERROR: A Blank Screen indicates low or depleted batteries.

SOLUTION: Remove and replace all 4 AA Batteries. Do not mix used and new batteries.

5]
ERROR MESSAGE: An (Er) Error Message indicates an abnormal memory function caused by a faulty internal component.

SOLUTION: Call Omron Customer Service, Toll Free, on (in the USA) 1-800-634-4850 (or on the equivalent Toll Free or Freefone Number in your country).

I hope this helps.

If this Answer goes some way towards helping to resolve your Query, Problem or Fault, please take the time to Rate it.
Doing so helps Solution Providers, like me, better understand the needs and the point of view of Askers, like you, and guides us in exploring the best ways to provide the most comprehensive, useful and high quality Responses and Solutions we can.

If you need further help, please do not hesitate to post another Question or add a Comment to this Question.

Thanks for using FixYa.

Sep 15, 2011 | Omron Health & Beauty

1 Answer

I WANT REMOVE OLD READINGS


...here's the simple solution from Omron themselves: just turn it on, hold the memory 'down' button until it starts scrolling thru all the past readings, then press n hold the start button until you see three dashes across all displays...this should take 2-5 seconds...enjoy!

Jul 19, 2011 | Omron Latest Model HEM-780 Automatic Blood...

1 Answer

More frequently my Omron HEM-780 seems flaky. It would apply pressure and count up to 112-120 then start deflating, then inflating, etc. Then it would count down by units and end up displaying E. It would...


Part of the problem may the the placement of the cuff. Generally, if it's not placed correctly. the cuff will re-inflate trying to find the correct stopping point for inflation. Any movement of your arm could cause this. Also, if you haven't replaced the batteries lately, you should do that. Also check to make sure there are no air leaks in the rubber tubing or the cuff. Check your Owners Manual for the exact placement of the cuff.

Feb 07, 2011 | Omron HEM 780 Automatic Blood Pressure...

2 Answers

I want to clear the memory on my omron hem 780


...nope, taking out the batteries won't help...here's the simple solution from Omron themselves: turn it on, hold the memory 'down' button until it starts scrolling thru all the past readings, then press n hold the start button until you see three dashes across all displays...this should take 2-5 seconds...enjoy!

Aug 20, 2010 | Omron Latest Model HEM-780 Automatic Blood...

2 Answers

Omron hem 780 how do I erase the memory


...here's the simple solution from Omron themselves: just turn it on, hold the memory 'down' button until it starts scrolling thru all the past readings, then press n hold the start button until you see three dashes across all displays...this should take 2-5 seconds...enjoy!

May 01, 2010 | Omron HEM 780 Automatic Blood Pressure...

1 Answer

Have Omron 780 Blood pressure monitor. 9 times out of 10 I get an E and no BP reading. What could be wrong?


It appears you may have a leak in the cuff. This will need to be verified by an authorised service facility. Contact OMRON they should be able to assist you.
CONSUMER SUPPORT TOLL FREE NUMBER: 1-877-216-1333

Customer Service Representatives are available
Monday through Friday - 8:30am to 4:30pm CT

Apr 20, 2010 | Omron HEM 780 Automatic Blood Pressure...

1 Answer

Cannot switch user


first take the blood pressure and once it has read your pressure then push the button to which user you want the pressure to remain on either A or B.

Jan 13, 2009 | Omron Latest Model HEM-780 Automatic Blood...

2 Answers

Reset the tru-read mod


I had a hard time setting this too.
You have to press the "set" button while the unit is OFF not on.


Dec 14, 2008 | Omron Latest Model HEM-780 Automatic Blood...

Not finding what you are looking for?
Omron Automatic Blood Pressure Monitor Options Large Adult Cuff Fits Arms 13 To 17 Logo

Related Topics:

38 people viewed this question

Ask a Question

Usually answered in minutes!

Top Omron Health & Beauty Experts

The Knight
The Knight

Level 3 Expert

74106 Answers

Maxumized

Level 2 Expert

70 Answers

Cindy Wells

Level 3 Expert

5100 Answers

Are you an Omron Health and Beauty Expert? Answer questions, earn points and help others

Answer questions

Manuals & User Guides

Loading...