I have a SmartLogic Life Brand Blood pressure monitor Model #5534. It wont work anymore.Either I start it and the cuff keeps inflating until I feel my arm is going to bust and I don't get a reading. Or I ger an ERR 3 message - but cuff is properly fitted and I have no leaks. I sometimes get ERR 1 message

Posted by on

Ad

6 Suggested Answers

6ya6ya
  • 2 Answers

SOURCE: I have freestanding Series 8 dishwasher. Lately during the filling cycle water hammer is occurring. How can this be resolved

Hi,
a 6ya expert can help you resolve that issue over the phone in a minute or two.
best thing about this new service is that you are never placed on hold and get to talk to real repairmen in the US.
the service is completely free and covers almost anything you can think of (from cars to computers, handyman, and even drones).
click here to download the app (for users in the US for now) and get all the help you need.
goodluck!

Posted on Jan 02, 2017

Ad
Max J
  • 22 Answers

SOURCE: WITH AUTOMATIC CUFF BLOOD PRESSURE MONITOR GETTING ERROR

E might represents insufficient inflation for your mom. For automatic inflation monitors usually there is a mechanism to keep inflating the cuff to higher pressures for people with higher BP. Check how is done. In some auto monitors,while cuff is inflated you keep pressing on START button until pressure reaches to some 15 to 20 mmHg higher than the pressure that auto usually stops and displays E.

Posted on Sep 08, 2009

Ad
DrRonJupiter
  • 10 Answers

SOURCE: mabis automaic blood pressure cuff, pressure monitor

Dear Edward: Likely, either cuff not tight enough or not in correct location on your arm or not holding device at correct height. Make sure you use Lithium Ion batteries. Try it on someone else. Dr Ron Grassi,Jupiter,Florida.USA. www.DrGrassi.com 30ad4e8.jpg

Posted on Oct 13, 2009

SEGlover
  • 1 Answer

SOURCE: Omron hem-704C blood pressure monitor show down

If your HEM 704C fell hard, it is probably the pressure transducer that is faulty, broken. To my experience it is cheaper to buy a new one then to have it repaired by a technician

Posted on Jan 19, 2010

Testimonial: "Sound advice."

jyackle5
  • 8197 Answers

SOURCE: looking for replacement arm cuff for my sunbeam

To get a replacement arm cuff call sunbeam at 1-800-394-2326

The cuff will cost you 12.95 there. If you need you can get the manual and order sheat on the link below. Page 10 of manual.

http://www.foodsaver.com/manuals/HAH/Manuals/7652.pdf

Good Luck

Posted on May 25, 2010

Testimonial: "i messed up you fix me thanks very much"

  • 1 Answer

SOURCE: The Sunbeam blood pressure kit doesn't inflate the cuff.

You have the same problem I do. The hose socket to the machine is broken.

Posted on Sep 06, 2010

Add Your Answer

Uploading: 0%

my-video-file.mp4

Complete. Click "Add" to insert your video. Add

×

Loading...
Loading...

Related Questions:

1 Answer

Cuff doesn't inflate lifesource ua767 bp monitor. How do I reset or fix this? Have brand new batteries.


Inflating a Blood Pressure Cuff - YouTube

? 1:01
www.youtube.com/watch?v=8e6qtnHa8ps
Sep 30, 2008 - Uploaded by expertvillage When you take a blood pressure reading it is important to inflatethe blood pressure cuff until you no longer ...

Sep 24, 2015 | Lifesource Health & Beauty

1 Answer

How do I fix a homedics BPA-201 Digital Blood Pressure Monitor. Or who do I send it to be fixed. It keeps displaying E1. But I checked all cuffs connections and they are fine.


Are you moving your arm at all after you press the start button. I just have a wrist cuff, but if I move anything after I've pushed the start button I get an error code.

Apr 02, 2015 | Homedics Health & Beauty

1 Answer

What does error 3 on my microlife blood pressure monitor


Repeat measurement keeping arm still. If inflation of the cuff takes too long, the cuff is not correctly seated or the hose connection's not tight. Check connections and repeat

Jan 10, 2015 | Microlife Health & Beauty

1 Answer

My blood pressure machine keeps giving me an error 3 it is a cvs brand


Several of the CVS brand blood pressure monitors are made by MicroLife USA. For those models, the Error 3 indicates that you may have been moving your arm while the unit was obtaining the readings. You need to repeat the measurement while staying still.

The other possible cause of the error are that the cuff didn't inflate in a reasonable length of time. Check that the cuff is properly placed on your arm. Also make sure that the inflation hose is tightly connected to the cuff socket on the unit.

I hope this helps.

Cindy Wells

Aug 14, 2014 | Health & Beauty

1 Answer

Error 3 is showing on my monitor ... What does it mean ?


On other blood pressure monitors from Life Source, the error 3 means that the blood pressure cuff isn't inflating properly. The cuff could be incorrectly fastened or the air connector is backwards or not connected to the cuff or the monitor. Check this on your monitor.

I hope this helps.

Cindy Wells

Jul 08, 2014 | Health & Beauty

1 Answer

The cuff squeezes my arm too much and it hurts. The reading is way too high everytime. I don't know how to adjust it.


do not inflate the cuff too much. if you know your usual blood pressure, just add about 5-10 mmHg when you inflate it so that you won't hurt yourself. for example, if your BP is usually 130/90, squeeze the cuff to 135-140 only. but if you doubt that your BP is way much higher than that by assessing on what you are feeling, just do not squeeze the cuff further than 200.

Jan 21, 2011 | Life Fitness Ultra Blood Pressure Monitor...

1 Answer

What does it mean when you get a reading then 3 consecutive readings say EE


What is the pattern of the EE? If the two E's are on the same line, then you didn't inflate the cuff to a sufficient pressure. (Wait 2-3 minutes before trying again and then inflate the cuff to 40 mmHg higher than the previous attempt.)

If the Es are arranged vertically, then you overinflated the cuff. Any [ressure above 300 mmHg is beyond the range that this bp monitor can handle. If you reached this point by getting several underinflation alerts, ask your medical professional to test your bp and recommend the starting inflation level that you need.

Make sure that you have the cuff on properly. You may need to consider a different bp pressure cuff size to get an accurate reading.

I hope this.

Cindy Wells

Sep 09, 2010 | Omron Model HEM 432C Manual Inflation...

1 Answer

It keep giving me a ee reading every time i try to use it


Hi. One of the most common error symbol that appears on this display panel is a EE with a down arrow and two numerals. The reason for the EE Error Indicator is usually caused because the cuff has been wrapped too loose. The cuff was under-inflated during the measurement. This error will also display if the cuff is worn and needs to be replaced. The EE symbol with a down arrow indicates the cuff is over-inflated. In most instances the cuff has been wrapped too tight or you are using the incorrect cuff size. Turn the monitor off. Read the instructions to correctly apply the cuff. Wait five to ten minutes then take another measurement. This should correct the issue.

Aug 19, 2010 | Omron Model HEM 432C Manual Inflation...

1 Answer

WITH AUTOMATIC CUFF BLOOD PRESSURE MONITOR GETTING ERROR


E might represents insufficient inflation for your mom. For automatic inflation monitors usually there is a mechanism to keep inflating the cuff to higher pressures for people with higher BP. Check how is done. In some auto monitors,while cuff is inflated you keep pressing on START button until pressure reaches to some 15 to 20 mmHg higher than the pressure that auto usually stops and displays E.

Mar 04, 2009 | Mabis Automatic Digital Blood Pressure...

1 Answer

When the pressure comes up on the monitor the top number was reading EE. It did it three times. I can't find the instructions. The machine is an Omron model: 741CREL. Thanks.


EE=Cuff under-inflated=not inflated enough. While cuff is inflated, keep pressing START button until pressure reaches 15-20 mmHg higher than the pressure when it stops and displays EE. Make sure at least 5 min allowed between readings.

Feb 21, 2009 | Omron Automatic Blood Pressure Monitor...

Not finding what you are looking for?
Health & Beauty Logo

Related Topics:

1,474 people viewed this question

Ask a Question

Usually answered in minutes!

Top Health & Beauty Experts

Robin Hill
Robin Hill

Level 3 Expert

1263 Answers

wcdoc1953
wcdoc1953

Level 2 Expert

278 Answers

yadayada
yadayada

Level 3 Expert

76109 Answers

Are you a Health and Beauty Expert? Answer questions, earn points and help others

Answer questions

Manuals & User Guides

Loading...