Question about Welch Allyn Medical

1 Answer

We have a Schiller AT-2 EKG machine that is not printing on the paper. Please let us know what to do.

Posted by on

1 Answer

  • Level 3:

    An expert who has achieved level 3 by getting 1000 points

    All-Star:

    An expert that got 10 achievements.

    MVP:

    An expert that got 5 achievements.

    Vice President:

    An expert whose answer got voted for 100 times.

  • Master
  • 532 Answers

Hello,

Does the printer work in auto mode ?

Did you put the paper correctly. There is a black mark on paper which is sensed by paper sensor .If you put paper up side down then paper sensor won't be able to sense any paper & gives you error for rifilling paper .

If this doesn't help you then You need to replace battery .

Thanks for using Fixya.

Posted on Jul 14, 2010

1 Suggested Answer

6ya6ya
  • 2 Answers

SOURCE: I have freestanding Series 8 dishwasher. Lately during the filling cycle water hammer is occurring. How can this be resolved

Hi,
a 6ya expert can help you resolve that issue over the phone in a minute or two.
best thing about this new service is that you are never placed on hold and get to talk to real repairmen in the US.
the service is completely free and covers almost anything you can think of (from cars to computers, handyman, and even drones).
click here to download the app (for users in the US for now) and get all the help you need.
goodluck!

Posted on Jan 02, 2017

Add Your Answer

Uploading: 0%

my-video-file.mp4

Complete. Click "Add" to insert your video. Add

×

Loading...
Loading...

Related Questions:

2 Answers

When auto print is pressed the machine blanks out and nothing prints


Try this link to slide share for a colour rendition of the user manual
http://www.slideshare.net/DiscountCardiology/schiller-at-2-ekg-manual

Aug 30, 2012 | Schiller Cardiovit At 2 Light Ecg Machine

1 Answer

At-2 lite model not printing visibly patient data and interpretation also have blurred upper part of EKG print. All leads are printed well. Is this setup problem or malfunction?


hi,
check the paper you have inserted. if the paper quality is bad then the printing can become bad. Use schiller original paper.

if again printing is bad then the thermal printing head of you machine is faulty or the. you have to replace the same.
ok

Apr 10, 2011 | Schiller Cardiovit At 2 Light Ecg Machine

1 Answer

The ECG machine expels the paper but with no ecg record on it, please advise?


hi,
Your machines thermal printing head have some problem. Thats why it doesn't print any thing on the paper. Not possible to repair thermal head. You have to replace it with a new one..
ok

Feb 17, 2011 | Schiller Welch Allyn Cardiovit At 2...

1 Answer

Schiller cardiovit at-2 we can only get 3\4 of a ekg it stops


hi,
ur Schiller at 2 has problem with its battery. just replace the same. its a12v battery.
also clean the paper sensor Whit an ear bud (do not use any chemical use only ear bud). u can see the paper sensor on the side of printer head. a square black portion.
ok

Jan 03, 2011 | Schiller Welch Allyn Cardiovit At 2...

1 Answer

It will print 3\4 of the ekg it stops !!


hi,
ur Schiller at 2 has problem with its battery. just replace the same. its a12v battery.
also clean the paper sensor With an ear bud (do not use any chemical use only ear bud). u can see the paper sensor on the side of printer head. a small square black portion.
ok

Jan 03, 2011 | Schiller Welch Allyn Cardiovit At 2...

2 Answers

Just ordered a schiller at1 ekg machine, just realized it is 3 channel, does it have a 1 channel mode where just one tracing will print?


hi,
u can make it single channel one by changing its setup. it will print single trace in the same paper. when u select this option it use more paper than the three trace.
ok

Dec 06, 2010 | Schiller Cardiovit At 1 Ecg Machine

3 Answers

Schiller AT-2 Plus, placed new paper in machine and machine keeps stating "out of paper" and will not print unless I hold down the manual print.


Your machine has the wrong paper in it or the paper sensor is not working properly.  There should be a black mark on the top left hand corner of each page.  The correct ekg paper and ekg repair is available from Medical Device Depot.

Feb 15, 2010 | Schiller Welch Allyn Cardiovit At 2...

2 Answers

The schiller at 2 plus ecg machine not printing ecg


Does it work in auto mode ? are you putting paper correctly means , there is a black mark on paper which is sensed by paper sensor .If you put paper up side down then paper sensor won't be able to sense any paper & gives you error for rifilling paper .If this doesn't help you then You need to replace battrey . Thanks

Jul 15, 2009 | Schiller Welch Allyn Cardiovit At 2...

1 Answer

SCHILLER AT 2 EKG ISN"T PRINTING ON AUTO PRINT I NEED TO KNOW HOW TO RESET IT. I'M LOOKING FOR A MANUAL.


my machine now printed 2 paper in auto print mode, i want this print one paper .Can you help me solve this problem?
hoanghung_hoanghung24@yahoo.com

Apr 07, 2009 | RTO Schiller AT-2i Electrocardiograph -...

Not finding what you are looking for?
Medical Logo

Related Topics:

44 people viewed this question

Ask a Question

Usually answered in minutes!

Top Welch Allyn Medical Experts

Electro Med Services...
Electro Med Services...

Level 3 Expert

6694 Answers

masud
masud

Level 3 Expert

572 Answers

Rhonda Webster
Rhonda Webster

Level 3 Expert

948 Answers

Are you a Welch Allyn Medical Expert? Answer questions, earn points and help others