Question about Health & Beauty

1 Answer

How do I adjust for monitor for machine to read higher BP

Posted by on

1 Answer

  • Level 3:

    An expert who has achieved level 3 by getting 1000 points

    All-Star:

    An expert that got 10 achievements.

    MVP:

    An expert that got 5 achievements.

    Master:

    An expert who has achieved Level 3.

  • Master
  • 604 Answers

Get to the hospital,if your bp is higher than the machine can read,youare having a heartattack

Posted on Apr 21, 2017

4 Suggested Answers

6ya6ya
  • 2 Answers

SOURCE: I have freestanding Series 8 dishwasher. Lately during the filling cycle water hammer is occurring. How can this be resolved

Hi,
a 6ya expert can help you resolve that issue over the phone in a minute or two.
best thing about this new service is that you are never placed on hold and get to talk to real repairmen in the US.
the service is completely free and covers almost anything you can think of (from cars to computers, handyman, and even drones).
click here to download the app (for users in the US for now) and get all the help you need.
goodluck!

Posted on Jan 02, 2017

udaikirun
  • 2 Answers

SOURCE: suddenly showing higher reading

the damage to sleeve might not b the exact reason all v need in a spigmomanometer is pressure . it might b a problem internally just chek that once with a biomedical engeneer if the meter is perfect immidiatly care ur health

Posted on Mar 18, 2009

Max J
  • 22 Answers

SOURCE: Irratic BP readings

Try 5 to 10 minutes between readings, all at the same conditions: seated on chair, lean back, feet on ground.....

Posted on Sep 07, 2009

  • 2 Answers

SOURCE: I keep getting low readings.

you should already be dead if your bp is 196/119.normal runs 120/80.hypertension starts at 140/90. wrist bp units are often incorrect. dump nurse and cuff. go see your doc or use the free units at kroger

Posted on Oct 07, 2010

Add Your Answer

Uploading: 0%

my-video-file.mp4

Complete. Click "Add" to insert your video. Add

×

Loading...
Loading...

Related Questions:

1 Answer

How do i adjust the start up pressure to 180?


If you move your arm or wrist during and after the cuff has inflated, the device will "bump" itself up to start reading at a higher number. Make sure you hold your wrist at about the same elevation as your heart, in order to get the correct blood pressure readings.

Sep 18, 2012 | Braun BP-5900 ExactFit Blood Pressure...

1 Answer

The cuff squeezes my arm too much and it hurts. The reading is way too high everytime. I don't know how to adjust it.


do not inflate the cuff too much. if you know your usual blood pressure, just add about 5-10 mmHg when you inflate it so that you won't hurt yourself. for example, if your BP is usually 130/90, squeeze the cuff to 135-140 only. but if you doubt that your BP is way much higher than that by assessing on what you are feeling, just do not squeeze the cuff further than 200.

Jan 21, 2011 | Life Fitness Ultra Blood Pressure Monitor...

1 Answer

What does it mean when you get a reading then 3 consecutive readings say EE


What is the pattern of the EE? If the two E's are on the same line, then you didn't inflate the cuff to a sufficient pressure. (Wait 2-3 minutes before trying again and then inflate the cuff to 40 mmHg higher than the previous attempt.)

If the Es are arranged vertically, then you overinflated the cuff. Any [ressure above 300 mmHg is beyond the range that this bp monitor can handle. If you reached this point by getting several underinflation alerts, ask your medical professional to test your bp and recommend the starting inflation level that you need.

Make sure that you have the cuff on properly. You may need to consider a different bp pressure cuff size to get an accurate reading.

I hope this.

Cindy Wells

Sep 09, 2010 | Omron Model HEM 432C Manual Inflation...

9 Answers

Where is omron bp monitor service center in india my bp monitor doesnot get enough preesure and shows no 50 on the screen.


call rehman on mobile no: +91-8754414889 for all your query of omron or other make of blood pressure monitor service problems

Jan 19, 2010 | Omron Automatic Blood Pressure Monitor...

1 Answer

Blood pressure reading 150/76. Manual Blood pressure 130/72


A monitor using Hg is only a valid monitor reading. Most electronic devices produce approximate results and are mostly good for taking several readings daily and obtaining a mean daily BP. Device calibration appears to be not accessible to the the user.Also, depending on age & condition of your dad, sequential reading must take place after elapse of enough time for circulation to be at steady state in the arm again, or if you immediately after a measurement make another, even with the same monitor you will obtain a different result.

Mar 23, 2009 | Health & Beauty

1 Answer

Suddenly showing higher reading


the damage to sleeve might not b the exact reason all v need in a spigmomanometer is pressure . it might b a problem internally just chek that once with a biomedical engeneer if the meter is perfect immidiatly care ur health

Mar 15, 2009 | Health & Beauty

2 Answers

Readings vary widely withing 5 mins. when repeated


You should not recheck BP until at least 15 minutes after the first check because it will always get higher on reading even with a manual BP machine. As a retired nurse I have both an automatic HoMedics which I I lost the manual for and a old type arm BP cuff. My automatic is pretty close to the other cuff in readings so I know it is close. Try waiting for a while before you recheck and not talk or move around when you take it.

Mar 10, 2009 | Health & Beauty

1 Answer

Irratic BP readings


Try 5 to 10 minutes between readings, all at the same conditions: seated on chair, lean back, feet on ground.....

Mar 04, 2009 | Sunbeam Digital Blood Pressure Monitor

1 Answer

WITH AUTOMATIC CUFF BLOOD PRESSURE MONITOR GETTING ERROR


E might represents insufficient inflation for your mom. For automatic inflation monitors usually there is a mechanism to keep inflating the cuff to higher pressures for people with higher BP. Check how is done. In some auto monitors,while cuff is inflated you keep pressing on START button until pressure reaches to some 15 to 20 mmHg higher than the pressure that auto usually stops and displays E.

Mar 04, 2009 | Mabis Automatic Digital Blood Pressure...

Not finding what you are looking for?
Health & Beauty Logo

Related Topics:

28 people viewed this question

Ask a Question

Usually answered in minutes!

Top Health & Beauty Experts

Robin Hill
Robin Hill

Level 3 Expert

1263 Answers

yadayada
yadayada

Level 3 Expert

75117 Answers

jeff tilburn
jeff tilburn

Level 3 Expert

862 Answers

Are you a Health and Beauty Expert? Answer questions, earn points and help others

Answer questions

Manuals & User Guides

Loading...