Question about Schiller Cardiovit At 1 Ecg Machine

Open Question

Schiller At-1 EKG machine won't print, filter light on

Posted by Anonymous on

6 Suggested Answers

6ya6ya
  • 2 Answers

SOURCE: I have freestanding Series 8 dishwasher. Lately during the filling cycle water hammer is occurring. How can this be resolved

Hi,
a 6ya expert can help you resolve that issue over the phone in a minute or two.
best thing about this new service is that you are never placed on hold and get to talk to real repairmen in the US.
the service is completely free and covers almost anything you can think of (from cars to computers, handyman, and even drones).
click here to download the app (for users in the US for now) and get all the help you need.
goodluck!

Posted on Jan 02, 2017

  • 2 Answers

SOURCE: Schiller Welch Allyn Cardiovit At 2

I HAVE SEEN THIS A FEW TIMES. I HAVE RELATED THIS TO THE PAPER CARRIES TRAY! CHECK THE TRAY AND MAKE SURE THE TIPS HAVE NOT BROKEN! THIS COMES FROM SOMEONE SLAPPING THE TRAY INTO PLACE INSTEAD OF COMPRESSING THE SMALL RELEASE.
MIKE
CALL
909-599-1789

Posted on Jan 29, 2009

  • 1 Answer

SOURCE: the window on our schiller AT 2 plus ekg machine is too dark

When you first turn on the machine, hit the up or down arrows to adjust the brightness of the LCD screen.

Posted on Mar 13, 2009

  • 1 Answer

SOURCE: Schiller AT-1 ecg not printing.

Turn on machine, wait a few seconds then press ALT-0-1-1

You will get a print out showing all the settings

Posted on Jul 19, 2009

  • 16 Answers

SOURCE: the schiller at 2 plus ecg machine not printing ecg

Does it work in auto mode ? are you putting paper correctly means , there is a black mark on paper which is sensed by paper sensor .If you put paper up side down then paper sensor won't be able to sense any paper & gives you error for rifilling paper .If this doesn't help you then You need to replace battrey . Thanks

Posted on Jul 22, 2009

SOURCE: need user manual for Welch Allyn EKG machine -

You can find a quick tutorial at http://www.youtube.com/watch_popup?v=Z8aaQvDqS48

Posted on Feb 21, 2010

Add Your Answer

Uploading: 0%

my-video-file.mp4

Complete. Click "Add" to insert your video. Add

×

Loading...
Loading...

Related Questions:

2 Answers

When auto print is pressed the machine blanks out and nothing prints


Try this link to slide share for a colour rendition of the user manual
http://www.slideshare.net/DiscountCardiology/schiller-at-2-ekg-manual

Aug 30, 2012 | Schiller Cardiovit At 2 Light Ecg Machine

1 Answer

The ekg machine light of filter is flickering and when printing only prints one line out of three appears as if no ink


HI,
Here i added a link which will help you to rectify the problem. Its seems to be the battery of your ECG machine is faulty.

Schiller AT 1 Support

Jun 25, 2012 | Schiller Welch Allyn Cardiovit At 2...

1 Answer

The Schiller AT-1 is not printing, when the patiente cable is conected the unit start to print and stop after 1 second and turn off. If I disconec the cable the unit keep printing


The Schiller AT1 is a battery powered unit. A common symptom of weak/defective battery's is that it will stop printing after 1-2 seconds of run. As to whether the patient cable effects this, I honestly don't know, It really shouldn't, baring some very weird internal short in the cable. The battery in the AT1 is beyond the average user to safely replace. Before anything else insure that the AC indicator light illuminates when the unit is plugged in. If not, you device might have a defective power supply. In either case, the unit needs repair in a Qualified shop.

Oct 28, 2011 | Schiller Welch Allyn Cardiovit At 2...

1 Answer

At-2 lite model not printing visibly patient data and interpretation also have blurred upper part of EKG print. All leads are printed well. Is this setup problem or malfunction?


hi,
check the paper you have inserted. if the paper quality is bad then the printing can become bad. Use schiller original paper.

if again printing is bad then the thermal printing head of you machine is faulty or the. you have to replace the same.
ok

Apr 10, 2011 | Schiller Cardiovit At 2 Light Ecg Machine

2 Answers

Just ordered a schiller at1 ekg machine, just realized it is 3 channel, does it have a 1 channel mode where just one tracing will print?


hi,
u can make it single channel one by changing its setup. it will print single trace in the same paper. when u select this option it use more paper than the three trace.
ok

Dec 06, 2010 | Schiller Cardiovit At 1 Ecg Machine

2 Answers

Schiller AT-1


check paper alignment
maybe papermark sensor problem also

Oct 16, 2009 | Schiller Cardiovit At 1 Ecg Machine

1 Answer

I ahve a schiller welch allen EKG machine, and now it only prints


If it stops printing completely in the middle of doing the print, it is because the battery is bad and needs to be replaced.
If it only prints some of the leads but it does print the full report it is because you have the leads mixed up (Wrong spot on the patient) or the leads are mislabeled or some of the clips or electrodes are bad.
I can fix these. Let me know if you need help.

Apr 05, 2009 | Schiller Welch Allyn Cardiovit At 2...

2 Answers

Schiller AT-1 ecg not printing.


Turn on machine, wait a few seconds then press ALT-0-1-1

You will get a print out showing all the settings

Feb 16, 2009 | Schiller Welch Allyn Cardiovit At 2...

Not finding what you are looking for?
Schiller Cardiovit At 1 Ecg Machine Logo

Related Topics:

122 people viewed this question

Ask a Question

Usually answered in minutes!

Top Schiller Medical Experts

Electro Med Services...
Electro Med Services...

Level 3 Expert

6694 Answers

The Knight
The Knight

Level 3 Expert

73591 Answers

Are you a Schiller Medical Expert? Answer questions, earn points and help others

Answer questions

Manuals & User Guides

Loading...