Wikipedia says: A centrifuge is a piece of equipment, generally driven by an electric motor (some older models were spun by hand), that puts an object in rotation around a fixed axis, applying a force perpendicular to the axis. A centrifuge is also used to separate the components of blood in blood banks.The centrifuge works using the sedimentation principle, where the centripetal acceleration causes denser substances to separate out along the radial direction (the bottom of the tube). By the same token lighter objects will tend to move to the top (of the tube; in the rotating picture, move to the centre).

Posted on May 01, 2013

1 Suggested Answer

6ya6ya
  • 2 Answers

SOURCE: I have freestanding Series 8 dishwasher. Lately during the filling cycle water hammer is occurring. How can this be resolved

Hi,
a 6ya expert can help you resolve that issue over the phone in a minute or two.
best thing about this new service is that you are never placed on hold and get to talk to real repairmen in the US.
the service is completely free and covers almost anything you can think of (from cars to computers, handyman, and even drones).
click here to download the app (for users in the US for now) and get all the help you need.
goodluck!

Posted on Jan 02, 2017

Add Your Answer

Uploading: 0%

my-video-file.mp4

Complete. Click "Add" to insert your video. Add

×

Loading...
Loading...

Related Questions:

1 Answer

Eppendorf Centrifuge 5430R lid does not open after pushing the 'Open' button in front display


Probably too late now but there is usually a small hole in the centrifuge where a thin shafts can be inserted to manually disengage the locking system.

Apr 19, 2014 | Science & Laboratory

1 Answer

5402 error 2


Dear Sir,

In a 5402 Eppendorf Centrifuge, Error 1 states primarily for an EEPROM error. EEPROM is a tiny memory chip that stores individual details of the centrifuge, mainly the microprocessor time base (for correct time set operations), the mains voltage (115 or 230 V, to inform and set the motor current levels), and the imbalance level adjustment for the imbalance sensor. When the centrifuge powers up, these data are readen, and if all OK, the centrifuge is ready to go. If one of them corrupts, especially the 115 or 230 Volt information, we get error 1 and no start allowed.
To check and reprogram this EEPROM we have to acess the service routine. This is done by by jumping together pins 1 and 5 of X8 conector, on the base of the electronic board, located behind the power cable panel in the back of the centrifuge. There are about 20 setting or check parameters, but usually 1 or 2 of the described above need to be reprogrammed. Please e-mail me for details: phdeletr@gmail.com.

Error 2 is Motor Tach (Speed sensor) error, caused usually if the cable pops off the electronic board. But the tach sensor itself, who stays in the bottom of the motor drive, could be broken or defective.

Regards,

Paulo H. F. Pimenta

Dec 13, 2013 | Eppendorf 5402 Refrigerated Centrifuge

2 Answers

Error 1 on the 5402 centrifuge how to repair problem


Dear Sir,

In a 5402 Eppendorf Centrifuge, Error 1 states primarily for an EEPROM error. EEPROM is a tiny memory chip that stores individual details of the centrifuge, mainly the microprocessor time base (for correct time set operations), the mains voltage (115 or 230 V, to inform and set the motor current levels), and the imbalance level adjustment for the imbalance sensor. When the centrifuge powers up, these data are readen, and if all OK, the centrifuge is ready to go. If one of them corrupts, especially the 115 or 230 Volt information, we get Error 1 and no start allowed.
To check and reprogram this EEPROM we have to acess the service routine. This is done by by jumping together pins 1 and 5 of X8 conector, on the base of the electronic board, located behind the power cable panel in the back of the centrifuge. There are about 20 setting or check parameters, but usually 1 or 2 of the described above need to be reprogrammed. Please e-mail me for details:

phdeletr@gmail.com.

Regards,

Paulo H. F. Pimenta

Dec 11, 2013 | Eppendorf 5402 Refrigerated Centrifuge

1 Answer

Show error 1


Dear Sir,

In a 5402 Eppendorf Centrifuge, Error 1 states primarily for an EEPROM error. EEPROM is a tiny memory chip that stores individual details of the centrifuge, mainly the microprocessor time base (for correct time set operations), the mains voltage (115 or 230 V, to inform and set the motor current levels), and the imbalance level adjustment for the imbalance sensor. When the centrifuge powers up, these data are readen, and if all OK, the centrifuge is ready to go. If one of them corrupts, especially the 115 or 230 Volt information, we get error 1 and no start allowed.
To check and reprogram this EEPROM we have to acess the service routine. This is done by by jumping together pins 1 and 5 of X8 conector, on the base of the electronic board, located behind the power cable panel in the back of the centrifuge. There are about 20 setting or check parameters, but usually 1 or 2 of the described above need to be reprogrammed. Please e-mail me for details: phdeletr@gmail.com.
Regards,

Paulo H. F. Pimenta

Dec 03, 2013 | Eppendorf 5402 Refrigerated Centrifuge

1 Answer

Manual/electrical


If Sorvall doesn't have it and an internet search doesn't help either you may be out of luck

Apr 25, 2013 | Sorvall RC-2B Centrifuge

1 Answer

My centrifuge 5702 rotor mentions error 3


The Eppendorf manual states that Error 3 means that there was an anomaly in the speed sensing system. The corrective action is to leave the power on with the centrifuge not spinning for several minutes and the error should clear itself.

If the error persists, then the speed sensor is probably damaged, and the unit will need to be serviced by a qualified technician.

Apr 16, 2013 | Eppendorf North America Eppendorf 5702...

1 Answer

S7 error message on Beckman Avanti J-E centrifuge


follow the instruction in the procedure in the last paragraph of thi link

http://www.labwrench.com/?community.posts/threadNo/962/subject/System-Error-S7/
but i think this will reset the rotor calibrtion or configuration. my system will stop the run on rotor ja-20 at 5000 rmp; and it change the rotor id set on ja-14 t 18 at 10000rpm.

Oct 30, 2012 | Beckman Coulter Avanti J-E...

1 Answer

How do you work the magnetic mixer base?


you place it where you place the centrifuge but you take the centrifuge off after that you use the magnetic capsule and put the beaker on it and then put the magnetic capsule in the beaker then lock the dome in place and turn it on

Jan 15, 2012 | Planet Toys Csi Dna Lab

Not finding what you are looking for?
Science & Laboratory Logo

Related Topics:

632 people viewed this question

Ask a Question

Usually answered in minutes!

Top Science & Laboratory Experts

William McMath

Level 2 Expert

178 Answers

kakima

Level 3 Expert

102366 Answers

Tom Chichester

Level 3 Expert

33322 Answers

Are you a Science and Laboratory Expert? Answer questions, earn points and help others

Answer questions

Manuals & User Guides

Loading...