Tip & How-To about Schiller Cardiovit At 1 Ecg Machine

How to set up

Check the video to know how to setup Schiller AT 1 and AT2.
Electronics Equipment Repair Center

Posted by on

Schiller Cardiovit At 1 Ecg Machine Logo

Related Topics:

Related Questions:

1 Answer

how do you set time and date on schiller at2 ekg machine


HI,
To set time on the schiller AT2 do the following.
1. Power ON the machine.
2. Press the following key sequence ALT,9,1,1.
3 Now enter the time using the digit key (HHMMSS)
You will hear the conformation beep indicating that the time is seted.

To set the date on schiller AT2 Do the following.

1. Press the following key sequence ALT ,9,2,2.
2. Enter the date using the digit key (DDMMYY)
You will hear a conformation beep indicating that the date is seted.
ok

Sep 08, 2011 | Schiller Welch Allyn Cardiovit At 2...

3 Answers

High, I have a Schiller AT-2 light that is printing too lightly suddenly. Hard to read the lower page tracing in particular. Had it 5 years, doing well. Anything to adjust that?


hi,
Your machines thermal printing head have some problem. Thats why it give printouts very light. Not possible to repair thermal head. You have to replace it.
ok

Feb 15, 2011 | Schiller Cardiovit At 2 Light Ecg Machine

1 Answer

change date time on Schiller AT2


Press ALT - 9 - 1 - 1. This starts SETUP. U should b abble 2 change date/time.

Mar 09, 2009 | Medical

Not finding what you are looking for?

426 people viewed this tip

Ask a Question

Usually answered in minutes!

Top Schiller Medical Experts

Electro Med Services...
Electro Med Services...

Level 3 Expert

6694 Answers

The Knight
The Knight

Level 3 Expert

74243 Answers

Are you a Schiller Medical Expert? Answer questions, earn points and help others

Answer questions

Loading...