20 Most Recent Schiller Cardiovit At 1 Ecg Machine Questions & Answers


Hi I have the same problem.. of you have already solved the problem please let me know

Schiller... | Answered on Apr 10, 2017


Check the leads. If any of the Leads is open the this problem can occur. You have to check the continuity of each cable for the lead end to socket. All leads should show a resistance of 10KOhm
If you wish to get some details About schiller AT1; check the site linked here. Pull up older posts.
http://electro-medical.blogspot.com/

Schiller... | Answered on Jun 13, 2016


hi,
You have to check two things. First you have to check the patient cable, and have to check is its electrodes are properly cleanded or not. if any one of the lead or the electrode have some problem then the machine will stop printing.
Second thing the battery inside the machine. its a 12V 2.3AH battery. If availabel try to replace the same. To replace the battery you have to oppen the machine.
ok

Schiller... | Answered on Apr 17, 2015

Tip

How to set up


Check the video to know how to setup Schiller AT 1 and AT2.
Electronics Equipment Repair Center

on Jan 31, 2015 | Schiller Cardiovit At 1 Ecg Machine

Tip

Part list


Part list and part number Schiller AT 2 ECG EKG.
Case, Upper, AT-2 (complete w / keybrd)
Case, Upper, AT-2+ (complete w / keybrd)
Case, Lower, AT-2 / 2+
Case, Lower, AT-2+ w / mem / tx

Label, Instruction, Clear Overlay AT-2 only
Label, Instruction, Clear Overlay AT-2+ only
Paper, AT-2/AT-2+, Tray Kit
Assy, Paper, Lid AT-2 / AT-2+
Assy, Paper, Rel, Catch, AT-2 / AT-2+
Assy, Paper Tray, Holds Paper AT-2 / AT-2+
Paper, Feed Roller, Black AT-2 / AT-2+
Assy, Thermal Print Head, AT-2 / AT-2+
Paper, Mark Detector AT-2 / AT-2+
Thermal, Print Head Springs (2-required)
Assy, Paper Drive Motor AT-2 / AT-2+



Electronics Equipment Repair Center

on Jan 31, 2015 | Schiller Welch Allyn Cardiovit At 2...


Bionet CardioCare 2000 Interpretive EKG for around $1200 is a good user friendly model...

below is a link to shop for new & used machines
http://www.cardiacdirect.com/index.php?page=products&id=1

I hope this is helpful.

Schiller... | Answered on May 27, 2013


hi,
You machine have complaint in its main board. Years back i got a machine with same complaint. After a lot of serch i found a track break it its main board, because of some sort of chemical reaction it the tracks. Search for the same in your machine. Just trace the two tracks from the power on switch. One end of it will be common connection for all the other switches And the other end of the power on switch will goto the power supply. As a multlayer pcb it will be very difficult to trace the track. Anyway just try it. Also check the SMD (surface ***** device) fuses in the power supply stage. Also check the power stage for any track break. If you are not able to do this then there is only one way contact the manufacture for support. It will be very expensive because they will tell you to replace the main board.
ok

Schiller... | Answered on Mar 18, 2013


Most likely needs a new battery. Can be ordered from R&D batteries.

Schiller... | Answered on Mar 18, 2013


HI,
check the paper u have inserted. it should be Schiller original paper with page marking. also clean the paper sensor. u can see the paper sensor near the side of printing head. a small square portion. clean it with an ear bud, don't use any solution for cleaning only ear bud.
ok

Schiller... | Answered on Mar 18, 2013


Problem in Battery, Replace it

Schiller... | Answered on Mar 18, 2013

Not finding what you are looking for?
Schiller Cardiovit At 1 Ecg Machine Logo

Popular Tags

43 questions posted

Ask a Question

Usually answered in minutes!

Top Schiller Medical Experts

Electro Med Services...
Electro Med Services...

Level 3 Expert

6694 Answers

The Knight
The Knight

Level 3 Expert

74473 Answers

Jim Goldsmith
Jim Goldsmith

Level 3 Expert

1889 Answers

Are you a Schiller Medical Expert? Answer questions, earn points and help others

Answer questions