20 Most Recent Welch Allyn Spot Vital Signs Monitor with NIBP and Stand Questions & Answers

0helpful
1answer

What is Error code E5.0 on the tempeature

SureTemp Plus error codes - Welch Allyn

https://www.welchallyn.com/content/dam/welchallyn/.../sap.../80016117LITPDF.pdf
nal code map. (ST+. HH). Details. Host controller action. 0. 0. No error - used for ACK ..... E5.0. Heater control error. If the problem persists, replace the module.
10/25/2018 5:34:38 AM • Welch Allyn Spot... • Answered on Oct 25, 2018
0helpful
1answer

Would like to know the

hi,
Here is it,
Spot, Spot LXi and VSM Mobile Stand With Basket SKU: 4700-60 The mobile stand for the Welch Allyn Spot, Spot LXi and VSM


You can check on the following link.

http://www.cardiacdirect.com/Product/Accessories/Spot-Vital-Signs-with-Blood-Pressure-and-Masimo-SpO2.html

OK
9/1/2011 3:14:47 PM • Welch Allyn Spot... • Answered on Sep 01, 2011
0helpful
1answer

We keep getting error codes

hi,
u didn't mensionend the error code number any where. error code number is must for find out the problem. any way do the following and inform me is it worked or no.

p { margin-bottom: 0.21cm; }

Power the Spot Vital Signs off, wait five seconds, and power on. If the error code

reappears then power the Spot Vital Signs off and disconnect the battery for five minutes.

Reconnect the battery and power on. Press the Blood Pressure Start/Stop button to reset flashing patient alarm conditions.

0k

12/30/2010 6:06:31 AM • Welch Allyn Spot... • Answered on Dec 30, 2010
0helpful
1answer

"Start" Button turns off vital signs monitor

Even if it is plugged in it will not operate correctly without a good battery. When the pump starts to take a blood pressure it draws a lot of power and this forces it to shut off. You should replace the battery and let it charge 12 hours and you should be in business.
If you have further questions you can contact me at [email protected] or through our website www.overstockme.com
7/18/2010 10:58:54 PM • Welch Allyn Spot... • Answered on Jul 18, 2010
0helpful
1answer

I am a trainee nurse Can you tell me how does

You just have to put the right size cuff on the patient and press the green button.  Here is the manual for the Welch Allyn Spot monitor http://www.welchallyn.com/documents/Patient%20Monitoring/Vital%20Signs%20Capture/Spot%20Vital%20Signs/usermanual_20070523_spotvs.pdf
2/21/2010 9:38:19 PM • Welch Allyn Spot... • Answered on Feb 21, 2010
0helpful
1answer

We find that it takes a long time for the bp cuff

Standard operation.
2/12/2010 4:21:44 AM • Welch Allyn Spot... • Answered on Feb 12, 2010
0helpful
1answer

What's error code 05?

it means it is unable to take blood pressure. Could be wrong size cuff, hole in cuff, cuff attached improperly, cuff not attached at all. if you need more assistance on this, call 8009894909 ask for Jeff or email at [email protected]
10/26/2009 4:47:01 PM • Welch Allyn Spot... • Answered on Oct 26, 2009
0helpful
1answer

Temperature

If you press the power button and the blood pressure button simultaneously, the monitor goes into programming mode. You hit the mode button until you reach the temperature and select the F or C by pressing the blood pressure button. To get out of programming you press the power button to power down and then power back up to use.
1/30/2009 1:56:30 PM • Welch Allyn Spot... • Answered on Jan 30, 2009
0helpful
1answer

Getting welch allyn spot vitals signs LXi machone to work

This link allows you to review error codes and gives suggested corrective action....

I hope this helps.

http://www.welchallyn.com/documents/Patient%20Monitoring/Vital%20Signs%20Capture/Spot%20Vital%20Signs%20LXi/usermanual_20080121_spotlxi.pdf
1/7/2009 8:43:55 AM • Welch Allyn Spot... • Answered on Jan 07, 2009
0helpful
1answer

Welch Allyn EKG CP100, on resetting, start up screen comes up but then 3 beeps and shuts off?.

only a guess .. sounds like you need to reload software from disc! or ifyiu know how. reset whilevpowered up . replace power and restart hopefully will restart without a reload .
10/30/2018 7:28:13 PM • Welch Allyn... • Answered on Oct 30, 2018
0helpful
1answer

We use a Welch Allyn

hi,
You have wrongly connected the limb leads. Limb leads are those leads which we connect in the arms and legs.
In your case the leads in the arm have been interchanged. Thats why you get the message and prints inverted avR and avR.
check the leads and properly place it.
ok
4/18/2011 3:55:56 AM • Welch Allyn... • Answered on Apr 18, 2011
0helpful
1answer

We have a Schiller AT-2

Hello,

Does the printer work in auto mode ?

Did you put the paper correctly. There is a black mark on paper which is sensed by paper sensor .If you put paper up side down then paper sensor won't be able to sense any paper & gives you error for rifilling paper .

If this doesn't help you then You need to replace battery .

Thanks for using Fixya.
7/14/2010 7:50:18 PM • Welch Allyn... • Answered on Jul 14, 2010
0helpful
1answer

I have a WelchAllyn/Schiller AT-2Plus

Sounds like the power supply unit inside has gone bad....you will have to have it replaced. Btw, I’m available to help over the phone in case u need at https://www.6ya.com/expert/sean_54c3a3dc48a7773c
4/6/2010 3:11:11 PM • Welch Allyn... • Answered on Apr 06, 2010
0helpful
1answer

I keep getting and alert that says "data error, please turn off" and doesn't work in my CardioCare 2000 ECG/EKG Machine. Can you help please?

Question edited for clarity and typo. Question moved from Cars and Trucks category. Too specialised for Fixya. This is professional medical equipment. Contact maker support or send to them for repair. No one reports a fix for that error message. https://www.google.com/search?q=cardiocare+2000+data+error+please+turn+off

CardioCare 2000 - Interpretive 12 Channel ...

Bionet America https://bionetus.com > Medical > ECGs/Spirometers The CardioCare 2000 is an economically priced interpretive 12 channel electrocardiogram (ECG/EKG) machine that does not compromise on performance.

CardioCare 2000 - Interpretive 12 Channel ...

Bionet America https://shop.bionetus.com > products > cardiocare-2000... CardioCare 2000 is an economically priced interpretive 12 channel electrocardiogram (ECG / EKG) machine that does not compromise on performance. US$995.00 ..
4/3/2024 3:43:34 PM • Medical • Answered on Apr 03, 2024
0helpful
1answer

Hartmann Tensoval Comfort Classic Blood Pressure Monitor. What does Error 2 mean please?

Question edited for a lot more clarity. Question moved from Heating and Cooling category The manual explains this very clearly. Here it is in case you lost it. The cuff wasn't fitted correctly. https://www.manualslib.com/manual/1242713/Hartmann-Tensoval-Comfort.html?page=141#manual ..
11/15/2023 4:32:12 AM • Medical • Answered on Nov 15, 2023
0helpful
1answer

I've got a 30 year old queen size Craftmatic Bed with a corded remote. It gradually stopped working. Any ideas?

Question edited for clarity. Question moved from Car Audio! It is nearly always abuse to the remote from bending the wire where is enters. to body, or the end where it enters the bed base. Craftmatic will repair it free of charge. http://crm.craftmatic.com/service/ ..
7/16/2023 4:35:44 PM • Medical • Answered on Jul 16, 2023
0helpful
1answer

OKIMAT 2SL adjustable bed. Where can I get a 5 pin power lead for it please?

Question edited for clarity. Question moved from Power Computers category. Amazon it seems https://www.google.com/search?q=okimat+2sl+5+pin+power+lead+replacement ..
4/29/2023 5:54:27 PM • Medical • Answered on Apr 29, 2023
0helpful
2answers

I need the User manual for Spirosift SP-5000 please?

some help here, but not really enough on Page 1. https://www.google.com/search?q=spirosift+sp-5000+manual ..
3/12/2023 6:00:32 AM • Medical • Answered on Mar 12, 2023
Not finding what you are looking for?
Welch Allyn Spot Vital Signs Monitor with NIBP and Stand Logo

19 questions posted

Ask a Question

Usually answered in minutes!

Top Welch Allyn Medical Experts

Sean Wright
Sean Wright

Level 3 Expert

2045 Answers

Brad Brown

Level 3 Expert

19187 Answers

ADMIN Andrew
ADMIN Andrew

Level 3 Expert

66839 Answers

Are you a Welch Allyn Medical Expert? Answer questions, earn points and help others

Answer questions

Manuals & User Guides

View Most Popular

Welch Allyn Medical

  • Welch Allyn Medical

Most Popular Question

temperature

  • Medical
Loading...